Simulation of Optimized One-Way Microvalves for Microfluidic Systems
Simulation of Optimized One-Way Microvalves for Microfluidic Systems
sourin fouladifard1
1) Iran University of Science and Technology Email:
IOI Code
:
ABEB-ABHFGC
Publication :
4th International Conference on Recent Advances in Engineering, Innovation and Technology - Belgium(eitconf.com)
Abstract :
Microfluidic systems rely on precise fluid control, where microvalves play a critical role in directing flow, preventing backflow, and enabling applications such as nanoparticle synthesis and biomolecule separation. This study employs Finite Element Analysis (FEA) via COMSOL Multiphysics to simulate four passive one- way microvalves—diffuser/nozzle, Tesla, sawtooth, and flap—focusing on their rectification performance under Reynolds numbers from 50 to 500. The simulations solve the Navier-Stokes equations for incompressible laminar flow, incorporating fluid-structure interactions to analyze flow velocity and rectification ratio. Results show the flap valve achieves the highest rectification ratio of 3.6, followed by sawtooth (1.5), Tesla (1.4), and diffuser/nozzle (1.12) at a Reynolds number of 490. The diffuser/nozzle valve with a 45° divergence angle exhibited optimal performance among its variants. These findings provide insights into designing efficient microvalves for biomedical devices, lab-on-a-chip technologies, and fluid manipulation systems.
Keywords : microfluidics microvalve rectification comsol