Validation method for the determination of Vardenafil in pharmaceutical dosage forms by RP-HPLC
Validation method for the determination of Vardenafil in pharmaceutical dosage forms by RP-HPLC
Mehdi Rezaei1
1) M.Sc in organic chemistry
IOI Code
:
XYCC-ABFZA
Publication :
International Conference on Science and Engineering(icesconf.com)
Abstract :
Abstract:
The HPLC method was found to be the simple, accurate, economic and rapid method for routine estimation of the Vardenafil in the tablets. The mobile phase consists of a methanol and purified water (65: 35 v/v). The flow rate of 1.0 ml / min is quite robust. The optimum wavelength for detection was found to be 220 nm which led to better detector response for the Vardenafil. The assay method was found to be linear in the concentration range of 32 to 63 µg/ml. The used concentration for quality control (QC) of the samples was 50 µg/ml. The slope and intercept value for calibration curve were y=12326x-264.14 (R2 =0.9946). The low RSD value (0.374 < 3%) indicates that this method is precise and accurate.
The mobile phase and flow rate selection was chosen based on the peak parameters (such as height, tailing, theoretical plates and capacity factor) and run times [1,2].
Keywords : vardenafil validation pharmaceutical